TSS HT sc Triclamp Suspended solids probe, stainless steel, for high temperatures, ball valve fitting

TSS HT sc Triclamp Suspended solids probe, stainless steel, for high temperatures, ball valve fitting
Produktnr.: LXV325.99.20001
Stykpris Kontakt os
Afsendes indenfor 2 uger

TSS sc probes have been specially developed for industrial applications

Measures both turbidity and suspended solids

8 measurement signals cover the total range and deliver measured values in conformity with the relevant standards

Excellent stability thanks to comprehensive compensation for interference factors

TSS sc has a unique compensation system to overcome the effects of air bubbles

Hvad er der i boksen

Turbidity & Suspended Solids sensor, user manual

Specifikationer

Applikation: High temperature
automatisk visker: Nej
Diameter: 40 mm
Flow: Max. 3 m/s (the presence of air bubbles affects the measurement)
Garanti: 24 måneder
Hvad er der i boksen: Turbidity & Suspended Solids sensor, user manual
installationsstil: Weld On TriClamp Fitting, High Temperature
Kabellængde: 10 m (optional extension cables available)
Kalibrering: Turbidity (TRB): Factory calibrated
Solids (TS): To be calibrated by customer on site
Zero point: Permanently calibrated in the factory
Kalibreringsmetode: Turbidity: Formazin or Stablcal Standard (at 800 NTU). Requires a calibration kit.
Suspended Solids: Sample specific, based on gravimetric analysis with a correction factor procedure.
Kompatibel med: SC200, SC1000, SC4500. All controllers sold separately
Længde: Inline sensor (TriClamp): 332 mm
Målemetode: Combined multiple beam alternating light method with infrared diode system and beam focusing
Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm
Solids (TS): 90º and 120º scattered light measurement, wavelength = 860 nm
Måleområde: Turbidity (TRB): 0.001 - 9999 NTU
Solids (TSS): 0.001 - 500 g/L
Materiale: Gasket: FKM; Wiper: PA (GF), TPV
Max Temperatur: 90 °C
Measuring range TSS: 0,001 - 500 g/L
Measuring range turbidity: 0,001 - 9999 FNU
Model: TSS HT sc
Monteringskonfigurationer: TriClamp
Nøjagtighed: Turbidity up to 1000 FNU/NTU: < 5% of the measured value ±0.01 NTU
Omgivelsestemperatur: 0 - 60 °C (briefly 80 °C)
  0 - 90 °C
Parameter: Turbidity, Suspended Solids
Repeterbarhed: TSS content: < 4 %
Turbidity: < 3 %
Responstid T90: 1 – 300 s justerbar
Specielle fordele: with Triclamp
Specielle notater: Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
> 10 cm (TSS)
Trykområde: < 10 bar or < 100 m
Vedligeholdelse interval: 1 h/måned
Vægt: 1,6 kg