Compliance:
Hazardous location, Maritime, CE
Drift:
0.03 pH per 24 hours, non-cumulative
Driftstemperatur :
-5 - 70 °C (23 - 158 °F) pHD and ORP
0 - 50 °C (32 - 122 °F) SS pHD
Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation)
Elektrode type:
General Purpose
Flow:
maks.
Flowhastighed:
3 m (10 ft.) per second, maximum
Følsomhed:
± 0,01 pH
Garanti:
24 måneder
Kabelforbindelse:
Digital
Kabellængde:
10 m PUR (polyurethane) 4-conductor with one shield, rated to 105°C
Kalibreringsmetode:
Two point automatic, one point automatic, two point manual, one point manual.
Kommunikation:
Modbus
Længde:
271.3 mm
Måleområde:
-2.0 to 14.0 pH
Materiale:
PEEK
Titanium
Monteringsmetode:
Insertion
Nøjagtighed:
± 0,02 pH
Opbevarings-betingelser:
4 - 70 °C, 0 - 95% relative humidity (non-condensing)
Repeterbarhed:
± 0.05 pH
Sensor cable:
10 m (33 ft.) polyurethane, 4-conductor cable with one shield, rated to 105°C (221°F)
Sensor gevind:
1" NPT
Temperaturføler:
NTC 300 Ω thermistor for automatic temperature compensation and analyser temperature readout
Temperaturkompensation:
Automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature,
additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °C
Temperatur nøjagtighed:
± 0.5 °C (± 0.9 °F)
Transmissions afstand:
1000 m (3280 ft.), maximum when used with a termination box.
100 m (328 ft.), maximum
Trykområde:
Maximum 10.7 bar . 6.9 bar for Digital Sensor at 70°C, and 6.9 bar for Analog Sensor at 105°C.
Vægt:
0,316 kg
Væskeberørte materialer:
PEEK or PPS, salt bridge of matching material with PVDF junction, glass process electrode, titanium ground electrode, and FKM/FPM O-ring seals (pH sensor with optional HF-resistant glass process electrode has 316 stainless steel ground electrode, and perfluoroelastomer wetted O-rings; consult factory for other available wetted O-ring materials)