EZ7663 Total Nitrogen + Total Phosphorus Analyser

EZ7663 Total Nitrogen + Total Phosphorus Analyser
Produktnr.: EZ7663.XXXXXXXX
Stykpris Kontakt os

Hvad er der i boksen

EZ7663 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)

Specifikationer

Afløb: Atmospheric pressure, vented, min. Ø 32 mm
Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analoge udgange: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal prøver pr. kanal: 1, 2, 4, or 8
automatisk rengøring: Ja
Beskyttelsesklasse: Analyser cabinet: IP55 / Panel PC: IP65
Certificeringer: CE compliant / ETL certified
Demineralised water: Consumption: 140 mL/analysis (hour)
Detektionsgrænse: TN: ≤ 2 mg/L
TP: ≤ 0.025 mg/L
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensioner (H x B x D): 690 mm x 465 mm x 330 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flowhastighed: 100 - 300 mL/min
Garanti: 24 måneder
Hvad er der i boksen: EZ7663 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferenser: TN: The main interferences are Br- and I-. When the amount of I- is 2.2 fold of the amount of TN, or the amount of Br- is 3.4 fold of the amount of TN, this will interfere on the test results. Dissolved organic matter, surfactants and Chromium (VI) interfere. Various inorganic substances not normally found in natural water, such as Chlorite [ClO₂-] and Chlorate [ClO₃-], may interfere.

TP: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V), Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Kalibrering: Automatic, 2-point; frequency freely programmable
Kuvetter: Standard measurement cycle time for both TN & TP: 60 minutes
Målemetode: TN: UV photometric measurement at 220 nm after persulphate digestion in alkaline medium, based on APHA 4500-NO3 (B)
TP: Colorimetric measurement at 700 nm using ascorbic acid reduction and molybdate colour solution after persulphate digestion in acidic medium, based on APHA 4500-P
Måleområde: 2 - 50 mg/L TN; 0.025 - 5 mg/L TP
Materiale: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Omgivelsestemperatur: 10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter: Total Nitrogen (TN), Total Phosphorus (TP)
Precision: Better than 3% (TN) and 2% (TP) full scale range for standard test solutions
Prøve kvalitet: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Prøvetemperatur: 10 - 30 °C
Reagent Requirements: Keep between 10 - 30 °C
strøm: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Tryk: By external overflow vessel
Validation: Automatic; frequency freely programmable
Vægt: 25 kg