Afløb:
Atmospheric pressure, vented, min. Ø 32 mm
Alarm:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analoge udgange:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal prøver pr. kanal:
1, 2, 4 or 8
Beskyttelsesklasse:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Certificeringer:
CE compliant / UL certified
Demineralised water:
For rinsing / dilution
Detektionsgrænse:
5 mg/L
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimensioner (H x B x D):
690 mm x 465 mm x 330 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flowhastighed:
100 - 300 mL/min
Garanti:
24 måneder
Hvad er der i boksen:
EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferenser:
Bromide, sulphide, iodide ions may interfere. Ferricyanide causes high results and must be removed. Chromate and dichromate interfere and should be reduced to chromic state or removed. Ferric iron interferes if present in an amount substantially higher than the amount of chloride. Chromic ion, ferrous ion and phosphate do not interfere. Fats, oil, proteins, surfactants and tar.
Kalibrering:
N.A.
Kuvetter:
10 - 15 minutes
Målemetode:
Potentiometric titration with silver nitrate (AgNO₃), conform with standard method APHA 4500-Cl (D)
Måleområde:
25 ‐ 500 mg/L
25 - 500 mg/L Cl-
Optional:
5 - 50 mg/L Cl-
5 - 125 mg/L Cl-
5 - 250 mg/L Cl-
Internal dispenser dilution (max. factor 100)
Materiale:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metodenavn:
Titrimetric
Model:
EZ4006
Omgivelsestemperatur:
10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter:
Chloride
Precision:
Better than 2% full scale range for standard test solutions
Prøve kvalitet:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Prøvetemperatur:
10 - 30 °C
Reagent Requirements:
Keep between 10 - 30 °C
strøm:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Strømforsyning:
100 - 240 VAC, 50/60 Hz
Tryk:
By external overflow vessel
Validation:
Automatic; frequency freely programmable
Vægt:
25 kg