Afløb:
Atmospheric pressure, vented, min. Ø 32 mm
Alarm:
1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analoge udgange:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal prøver pr. kanal:
1, 2, 4 or 8
Beskyttelsesklasse:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Certificeringer:
CE compliant / UL certified
Demineralised water:
For rinsing / dilution
Detektionsgrænse:
≤ 10 mg/L
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimensioner (H x B x D):
690 mm x 465 mm x 330 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flowhastighed:
100 - 300 mL/min
Garanti:
24 måneder
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferenser:
Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Kalibrering:
Automatic, 2-point; frequency freely programmable
Kuvetter:
10 - 15 minutes
Målemetode:
Acid-base titration with hydrochloric acid
Måleområde:
100 - 5000 mg/L CaCO3
Optional:
10 - 500 mg/L CaCO3
25 - 1250 mg/L CaCO3
50 - 2500 mg/L CaCO3
Internal dispenser dilution (max. factor 100)
100 - 5000 mg/L CaCO₃
Materiale:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metodenavn:
Titrimetric
Omgivelsestemperatur:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter:
Alkalinity, free
Precision:
Better than 2% full scale range for standard test solutions
Prøve kvalitet:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Prøvetemperatur:
10 - 30 °C
Reagent Requirements:
Keep between 10 - 30 °C
strøm:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Tryk:
By external overflow vessel
Validation:
Automatic; frequency freely programmable
Vægt:
25 kg