EZ3507 Fluoride Analyser

EZ3507 Fluoride Analyser
Produktnr.: EZ3507.XXXXXXXX
Stykpris Kontakt os

Hvad er der i boksen

EZ3507 Fluoride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Fluoride Electrode, combined, 1 x empty 5L Reagent Container with Fittings (for F Standard Solution) and 1 x empty 10L Reagent Container with Fittings (for Buffer Solution)

Specifikationer

Afløb: Atmospheric pressure, vented, min. Ø 32 mm
Alarm: 1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analoge udgange: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal prøver pr. kanal: 1, 2, 4 or 8
automatisk rengøring: Ja
Beskyttelsesklasse: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Certificeringer: CE compliant / UL certified
Demineralised water: For rinsing
Detektionsgrænse: ≤ 0.1 mg/L
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensioner (H x B x D): 690 mm x 465 mm x 330 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flowhastighed: 100 - 300 mL/min
Garanti: 24 måneder
Hvad er der i boksen: EZ3507 Fluoride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Fluoride Electrode, combined, 1 x empty 5L Reagent Container with Fittings (for F Standard Solution) and 1 x empty 10L Reagent Container with Fittings (for Buffer Solution)
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferenser: Metal ions like aluminium > 72 mg/L, calcium > 108 mg/L and iron > 150 mg/L. Fats, oil, proteins, surfactants and tar.
Kalibrering: Automatic, 2-point; frequency freely programmable
Kuvetter: 8 minutes
Målemetode: Discontinuous measurement by combined ion-selective electrode with standard addition
Måleområde: 0.5 - 10 mg/L F-

Optional:
0.1 - 2.5 mg/L F-
0.25 - 5 mg/L F-
  0.5 - 10 mg/L F⁻
Materiale: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Model: EZ3507
Omgivelsestemperatur: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter: Fluoride
Precision: Better than 2% full scale range for standard test solutions
Prøve kvalitet: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Prøvetemperatur: 10 - 30 °C
Reagent Requirements: Keep between 10 - 30 °C
strøm: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Strømforsyning: 100 - 240 VAC, 50/60 Hz
Tryk: By external overflow vessel
Validation: Automatic; frequency freely programmable
Vægt: 25 kg