EZ2300 Total Aluminium + Aluminium Al(III) Analyser

EZ2300 Total Aluminium + Aluminium Al(III) Analyser
Produktnr.: EZ2300.XXXXXXXX
Stykpris Kontakt os

Hvad er der i boksen

EZ2300 Aluminum Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 4 x empty 2.5L Reagent Containers with Fittings (for Acid, Buffer, Colour & Mixed Reagent Solution)

Specifikationer

Afløb: Atmospheric pressure, vented, min. Ø 32 mm
Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analoge udgange: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Antal prøver pr. kanal: 1, 2, 4 or 8
Beskyttelsesklasse: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Certificeringer: CE compliant / UL certified
Demineralised water: For rinsing / dilution
Detektionsgrænse: ≤ 10 µg/L
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensioner (H x B x D): 690 mm x 465 mm x 330 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flowhastighed: 100 - 300 mL/min
Garanti: 24 måneder
Hvad er der i boksen: EZ2300 Aluminum Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 4 x empty 2.5L Reagent Containers with Fittings (for Acid, Buffer, Colour & Mixed Reagent Solution)
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferenser: Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Kalibrering: Automatic, 2-point; frequency freely programmable
Kuvetter: 20 min Total Al (dilution + 5 min)
30 min Total Al & Al (III)
Målemetode: Colorimetric measurement using pyrocatechol violet method at 578 nm
Måleområde: 0 - 150 µg/L Al
  10 - 150 µg/L Al

Optional:
10 - 75 µg/L
80 - 600 µg/L (with internal dilution)
160 - 1500 µg/L (with internal dilution)
Materiale: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Model: EZ2300
Omgivelsestemperatur: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Parameter: Aluminum, total; Al(III), dissolved
Precision: Better than 2% full scale range for standard test solutions
Prøvetemperatur: 10 - 30 °C
Reagent Requirements: Keep between 10 - 30 °C
strøm: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Tryk: By external overflow vessel
Validation: Automatic; frequency freely programmable
Vægt: 25 kg